15A2135L 24
Main Menu
Pipette Calibration?
Press NO or MENU SCROLL.
Main Menu
Instr. Calibration?
Press YES. The following prompt will be displayed:
Instrument Cal. Menu
Perform Instr. Cal.?
Press YES to continue with an instrument calibration.
The instrument will then check the temperature, and if it is outside
the specified range for proper operation (18 °C to 28 °C), the
following prompt will be displayed:
Temp. out of range
Continue? (Y/N)
If NO is pressed, the following prompt will be displayed:
Main Menu
Pipette Calibration?
If YES is pressed to continue with the calibration, there is no
certainty that the results the PCS
obtains will meet the system
specifications.
5.2.2 At the following prompt:
Enter Operator ID:
__
Enter a number (not exceeding 20 characters), identifying the
operator for this Instrument Calibration. For alphabetic character
entry, see Appendix E.