Universal DrillSaw Battery Charger User’s Guide
DFU-0256-1 Rev. 0 Page 24 of 35
6
6
.
.
0
0
Cleaning and Disinfecting
6.1 Charger AR-400UBC
The AR-400UBC charger is provided non-sterile and should not be sterilized.
The AR-400UBC charger can be cleaned/disinfected using a cloth and commercially
available surfactants or surface disinfectants only. The AR-400UBC must not be submersed
in any liquid.
Always place the Main Power switch to the “Off position” and disconnect the power before
cleaning the AR-400UBC charger.
Always comply with the instructions issued by the manufacturer of the
cleaning and/or disinfectant regarding concentration, exposure times,
temperature and material compatibility.
NEVER clean the charging contacts with liquid cleaners. If there is
dust, remove with dry compressed air.
DO NOT clean the device with abrasive cleaning or disinfectant
compounds, solvents, or other materials that could scratch or damage
the device.
DO NOT clean the device without a cover or charging cup present on
the charging bays.