SECTION 4 PREPARATION FOR USE AND INSTALLATION
_______________________________________________________________________
4-5
• Verify that light to right of ON/Standby button is yellow.
3. Start up monitor by pressing the ON/Standby button (lower right corner).
• Verify that light to right of ON/Standby button is green.
• Verify all self-tests complete successfully. A beep tone sounds.
• Verify next screen says “Connect BISx.”
4. Connect BISx to monitor. Connect PIC and sensor to BISx.
• Verify screen says, “BISx Initialization Complete.”
• Verify SENSOR CHECK begins.
5. Disconnect power cord from rear of monitor.
• Verify ‘OPERATING ON BATTERY BACKUP’ is displayed.
• Verify battery icon displays below BIS number.
6. Reconnect power cord.
• Verify battery icon is not displayed below BIS number.
• Verify “OPERATING ON BATTERY BACKUP” is not displayed.
7. End of install.