SECTION 8 SERVICING THE BIS VISTA MONITORING SYSTEM
8-9
8. Set the screws in the holes. Slowly tighten to 45 in-oz.
8.8 Replacing the BISx Bulkhead Connector
Caution:
All repairs to the BIS VISTA Monitoring System should be made only
by a qualified Biomedical Engineering Technician or other authorized
personnel.
Use only the parts and tools specified. Use of any others may damage
the instrument.
8.8.1 Parts and Tools Required
Parts Required: Tools Required:
BISx Bulkhead Replacement Kit P/N 195-0052
includes:
Torque screwdriver with Torx bit size
T-7.
BISx Bulkhead,
P/N 175-0052
Gasket, P/N 150-0187
Torx head screws, P/N 606-0008