SECTION 7 DIAGNOSTICS AND TROUBLESHOOTING
________________________________________________________________________
7-4
7.3 BIS VISTA System Messages and Corrective
Actions
When an alarm is triggered, a message appears in the Message Region of the screen. Possible
messages and the recommended operator actions are listed below:
BIS VISTA Messages and Operator Actions
Status Messages: Causes: Corrective Actions:
01
Connect BISx
1. BISx disconnected.
2. Defective BISx cable.
3. Defective BISx.
4. Defective monitor.
1. Connect BISx. Verify all
cable connections.
2. Replace Monitor Interface
Cable.
3. Replace the BISx.
4. Replace monitor.
13
Re-prep Sensor
1. Incorrect sensor
application.
2. Poor sensor
connections.
3. Sensor Check fails
4. Defective PIC.
5. Defective BISx.
1. Read Instructions on
sensor package and re-prep
sensor.
2. Check sensor connections.
3. Re-prep again or replace
sensor. Verify Sensor
Check passes.
4. Replace the PIC.
5. Replace BISx Bulkhead
Connector or replace BISx.
14
Sensor Disconnected
1. Disconnected sensor.
2. Poor or contaminated
connection between
sensor and PIC.
3. Disconnected PIC.
4. Defective PIC.
5. Defective BISx.
1. Connect the Sensor.
2. Connect/clean connection
between sensor and PIC.
3. Connect the PIC.
4. Replace the PIC.
5. Replace BISx Bulkhead
Connector or replace BISx.