237
Base Accuracy
DC: ±(2% of reading + 2uA)
15Hz < f < 100KHz : ±(10% of reading
+ 2uA)
100KHz < f < 1MHZ : ±(10% of reading
+ 2uA)
Range 5
504uA ~ 999.9uA, 1000uA ~ 8400uA,
frequency 15Hz - 1MHz
Resolution
1uA
Base Accuracy
DC: ±(2% of reading + 2uA)
15Hz < f < 100KHz : ±(10% of reading
+ 2uA)
100KHz < f < 1MHZ : ±(10% of reading
+ 2uA)
Range 6
1200uA ~ 8399uA, 8.40mA ~
10.00mA, frequency 15Hz
- 1MHz
Resolution
1uA, 0.01mA
Base Accuracy
DC: ±(2% of reading + 3counts)
,15Hz < f <100KHz: ±(10% of reading
+ 2counts)
DISPLAY FIXED RANGE (<6% OF RANGE)
Touch Current Display (Peak)
(AC+DC)
Range 1
N/A
Resolution
N/A
Additional Error
N/A
Range 2
0.0uA ~ 7.8uA, frequency 15Hz
- 1MHz
Resolution
0.1uA
Additional Error
DC: add ±(2% of reading + 0.2% of
range) to base accuracy
15Hz < f < 100KHz : add ±(2% of
reading + 0.2% of range) to base
accuracy
100KHz < f < 1MHZ : add ±(2% of
reading + 0.5% of range) to base
accuracy
Range 3
0.0uA ~ 31.5uA, frequency
15Hz - 1MHz
Resolution
0.1uA
Additional Error
DC: add ±(2% of reading + 0.2% of
range) to base accuracy
15Hz < f < 100KHz : add ±(2% of
reading + 0.2% of range) to base
accuracy
100KHz < f < 1MHZ : add ±(2% of
reading + 0.5% of range) to base
accuracy

Table of Contents