EasyManua.ls Logo

Associated Research OMNIA II 8257 - DISPLAY AUTO RANGE

Default Icon
264 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
241
Range 5
0.0uA ~ 504uA, frequency 15Hz
- 1MHz
Resolution
1uA
Additional Error
15Hz < f < 100KHz : add ±(2% of
reading + 0.2% of range) to base
accuracy
100KHz < f < 1MHZ : add ±(2% of
reading + 0.5% of range) to base
accuracy
Range 6
0.0uA ~ 600uA, frequency DC
100KHz
Resolution
1uA
Additional Error
15Hz < f < 100KHz : add ±(2% of
reading + 0.2% of range) to base
accuracy
DISPLAY AUTO RANGE
Touch Voltage Display (RMS)
(AC+DC)
Range 1
0.0mV ~ 32.0mV, frequency DC,
15Hz 1MHz
Resolution
0.1mV
Accuracy
DC , 15Hz < f <100KHz: ±(2% of
reading + 3counts)
100KHz < f < 1MHZ : ±5% of reading
(10.0mV - 999.9mV)
Range 2
28.0mV ~ 130.0mV, frequency DC,
15Hz - 1MHz
Resolution
0.1mV
Accuracy
DC , 15Hz < f <100KHz: ±(2% of
reading + 3counts)
100KHz < f < 1MHZ : ±5% of reading
(10.0mV - 999.9mV)
Range 3
120.0mV ~ 525.0mV, frequency DC,
15Hz - 1MHz
Resolution
0.1mV
Accuracy
DC , 15Hz < f <100KHz: ±(2% of
reading + 3counts)
100KHz < f < 1MHZ : ±5% of reading
(10.0mV - 999.9mV)
Range 4
400mV ~ 999.9mV,1000mV ~
2100mV, frequency DC, 15Hz - 1MHz
Resolution
0.1mV, 1mV
Accuracy
DC , 15Hz < f <100KHz: ±(2% of
reading + 3counts)
100KHz < f < 1MHZ : ±5% of reading
(10mV - 8500mV)
Range 5
1800mV ~ 8400mV, frequency DC,
15Hz - 1MHz
Resolution
1mV

Table of Contents