FG_D6_PROFIBUS_EN_01 / 2020-06-17
Details writing table structure
Bit 0 = 1 > Reset (stop the current cycle).
Bit 1 = 1 > Start (starting a test cycle).
Bit 2 = 1 > Special cycle (start a special cycle, example: regulator adjust).
Bit 3 = 1 > Program selection.
Bit 4 = 1 > Read the FIFO cycles results (the FIFO contains the 8 lasts results,
standard mode only).
Bit 5 = 1 > Read of the parameters.
Bit 6 = 1 > Write of the parameters.
Bit 7 = 1 > Reset of the results FIFO (reset all available results in the FIFO).
Bit 0 = 1 > Read of the instrument configuration.
Bit 1 = 1 > Read of the configuration / extended menu bits.
Bit 2 = 1 > Read of the function bits.
Bit 3 = 1 > Write of the configuration / extended menu bits.
Bit 4 = 1 > Write of the function bits.
Bit 5 = 1 > Read of the program name.
Bit 6 = 1 > Write of the program name.
Bit 7 = 1 > Read last result.
Address 06h: Number of the program to be selected.
Address 07h = 0.
Address 08h: Special cycle. Address 09h=0.