4
1.2 Intended use
Name: ATMOS C 361
Main functions: Suctionofsecretion,rinsinguidsandtem-
porarilycollectionofbodyuids.
Med. indications / application: For surgeries e.g. suction of
wound cavities, abscesses, etc.
Inendoscopyforthesuctionofsecretionorrinsinguids.
Forspontaneoussuctionofbodyuids.
Specication of the main function: Drainage and temporary
collectionofbodyuids.Anelectricsuctionpumpisusedto
generate negative pressure. An additional secretion canister
must be attached to allow for temporary collection of drained
bodyuids.
Application organ: Naturalorices and openingsresulting
from a surgical intervention (entire body; human and animals).
Application time:
Short-term use on the patient (< 30 days)
Application site: The application site is the clinical, outpatient,
medicalpracticeandveterinaryeld.Thedevicemayonlybe
usedbystawhohavebeenmedicallytrainedandinstructed.
Contraindications:
No application in low-vacuum range e.g.thoracic and wound
drainage.
Noapplicationoutsideofmedicalelds.
Nosuctionofammable,corrosiveorexplosiveuids/gases
Not suitable for use as a vacuum extraction system.
The product is: Xactive □notactive
Sterility: Not necessary
Single-use product / reprocessing:
The device and parts of the accessories are reusable. For
information on reprocessing, cleaning and disinfection please
see the operating instructions.
1.3 Function
● The ATMOS C 361 is a mains-operated surgical suction
device, the core of which is a high-performance dia-
phragm-type pump. The pump generates a vacuum inside
the hose and secretion canister system, allowing secretions
to be withdrawn and collected. Using a vacuum regulator
and the vacuum gauge, the target vacuum and thus the
suction capacity can be precisely adjusted.
● Severalsecretioncanistersofdierentsizesareavailable
for use with the secretion canister system (chapter 9.0
Spare parts and accessories). A hydrophobic DDS bacterial
andvirallterinthesecretioncanisterlidisimplementedto
prevent that secretion is sucked into the pump or bacteria
and viruses enter the inside of the device.
● A trolley with standard rail is available for mobile use.
1.0 Introduction