Operating instructions
ATMOS C 161 Aspirator
ATMOS C 261 Aspirator
ATMOS C 161 Battery
English
2022-07 Index 01
GA1GB.310105.0

Table of Contents

Questions and Answers

  • B
    ballardjenniferAug 11, 2025
    What to do if Atmos Medical Equipment device does not start?
    • R
      Rachel BatesAug 11, 2025
      If your Atmos Medical Equipment device isn't starting, there are a few things you can check: * **Battery:** Ensure the battery isn't discharged by connecting the recharging unit power supply for 1 to 2 hours. * **Plug Connections:** Verify that the recharging unit power plug and the power cable are securely seated. The indicator lamp should light up when the device is correctly connected.
  • E
    elopezAug 16, 2025
    Why is high vacuum indicated on Atmos Medical Equipment?
    • C
      coxrichardAug 16, 2025
      A high vacuum indication on your Atmos Medical Equipment could be due to: * A blocked hydrophobic DDS bacterial and viral filter, in which case you should replace it. * The overflow protection float blocking the suction lid inlet. Check the suction lid inlet; if necessary, empty the rinsing canister, clean the oversuction protection, and ensure the float ball can move freely.
  • M
    Megan HolmesAug 22, 2025
    What to do if Atmos Medical Equipment switches off automatically when switched on?
    • S
      Steven TurnerAug 22, 2025
      If your Atmos Medical Equipment switches off automatically after being turned on, it might be due to applied vacuum (over 40%). Ventilate the vacuum; after that, start-up should be possible again.
  • J
    Jesse MercerAug 29, 2025
    What to do if low or no vacuum is displayed on Atmos Medical Equipment?
    • K
      Kelly NguyenAug 29, 2025
      If your Atmos Medical Equipment shows low or no vacuum, consider these potential causes: * **Filter:** Ensure the hydrophobic DDS bacterial and viral filter is correctly inserted. * **Leaks:** Check that the suction lid and hose are tightly fitted. * **Valve Plates:** If secretion or blood has been sucked in, causing the unit’s valve plates to stick together, remove the hydrophobic DDS bacterial and viral filter and place it back on the connecting pieces, ensuring it is seated securely. * **Suction lid:** Check suction lid is fitted tightly. * **Oversuction:** In case of oversuction, the device must be sent in for repair.
  • D
    donald81Sep 3, 2025
    How to fix insufficient capacity in Atmos Medical Equipment?
    • A
      Alexandria MitchellSep 3, 2025
      Insufficient capacity in your Atmos Medical Equipment may arise from: * A discharged battery. In this case, charge the battery. * A blocked bacterial and viral filter. If the hydrophobic DDS bacterial and viral filter is blocked, exchange it.