24
Operation
4.0 Operation
Before following these instructions for use, please read the preceding chapter of
your respective version of the ATMOS C 161 Aspirator, ATMOS C 261 Aspirator or
ATMOS C 161 Battery!
WARNING
Ensure that the device is always functional and ready for use!
Thepatientcouldsuocate.
• Suction procedures in the respiratory tract must only be implemented following
appropriatebriengfromtrainedandinstructedmedicalprofessionals.
• Make sure that the secretion canister is emptied in good time. As soon as the
canister is half-full, it must be emptied (this applies to all areas of application)!
• Iftheliquidlevelistoohigh,theoverowprotectionistriggeredandthesuction
stops. Empty the canister.
• Check the vacuum display regularly!
• If secretion has been sucked into the pump due to improper use or tampering, the
device must be repaired by ATMOS or a service authorised by ATMOS.
• Use appropriate suction catheters, hose-rinsing apertures or suction instruments.
• Observe the liquid level in the secretion canister while suctioning.
WARNING
Risk of infection by lack of hygiene or damaged components!
Risk of deadly diseases being transmitted.
• Alwaysusenewconsumablesforeverypatient(DDSbacterialandvirallter
cartridge and suction hose).
• Always check whether hoses or the secretion canister system are damaged before
using the device. Replace any damaged parts.
WARNING
Electric shock due to damaged device!
Cardiac arrhythmias may be caused.
• Always check the device and mains power unit for damage before use.
• Replace any damaged parts immediately.
• Do not use the device if it is damaged.
Ambient conditions during operation
• Temperature: +10 to +35 °C
• Relative humidity: 20 to 80 %
• Air pressure 700 to 1060 hPa
4.1 Switching on the device
& The device should only be left on for as long as you need it. This will prolong the
battery life (for ATMOS C 161 Battery).
1. Presstheon/obuttontoswitchonthedevice.
» The pump starts.
4.2 Switching off the device
1. Switchothedevicebypressingtheon/obutton.