Cleaning/disinfection
5.1.8 Canister lid
Prior to using the device on a new patient, the canister lid must always be disinfected
using an instrument disinfectant as recommended on Page 38. Prior to disinfection,
pleaseensurethattheDDSbacterialandvirallterhasbeenremovedandthecanister
liddismantledintoitsindividualparts(lid,oatballandhoseconnector).
& Inordertoachieveabettercleaningeect,rinsetheindividualpartsunderclear
water for at least 10 seconds. Once this is done, disinfect the individual parts.
& Please observe the operating instructions for the respective disinfectant.
& For instructions on how to remove the canister lid, see chapter 4.7.2 on page 27.
& If the device is still being used by the same patient, we recommend rinsing the
canister lid and its individual parts under clear water after every suction procedure.
ATTENTION
RemovetheDDSbacterialandvirallterbeforedoingthis!
5.1.9 Device surface
Prior to using the device on a new patient, the entire surface of the device must
always be cleaned with a damp (not wet) cloth and disinfected using one of the
surface disinfectants listed in “5.5 Recommended surface disinfectants” on page 39.
If the device is still being used by the same patient, the device surface should still be
cleaned with a damp (not wet) cloth at least once a week and then disinfected using
one of the surface disinfectants listed on “5.5 Recommended surface disinfectants”
on page 39.
& Somedisinfectantsmaycausediscolourationofthematerial.Thisdoesnotaect
the functionality of the device.
& Always ensure that the device is disconnected from the mains power before com-
mencing cleaning!
& The device should never be autoclaved, rinsed under running water or immersed
into any liquids!
5.1.10 Accessories
Hose reel (REF 313.0007.0)
Trolley (REF 320. 0070.0)
Device support (REF 313.0012.0)
& Prior to using the device on a new patient, the entire surface of the hose reel, the
trolley and the device support must always be cleaned with a damp (not wet) cloth
and then disinfected using one of the surface disinfectants listed in “5.5 Recom-
mended surface disinfectants” on page 39.