16
For your safety
2.0 For your safety
The safety of the ATMOS C 161 Aspirator, ATMOS C 261 Aspirator and ATMOS C 161
Battery complies with all generally accepted engineering practices and the provisions of
the German Medical Devices Act.
2.1 General safety instructions
Report all serious incidents that have occurred in connection with this product to the
manufacturer and your national competent authority.
Familiarise yourself with the device in good time so that you are capable of operating it
at any time.
Only a fully functional product meets the safety requirements of users, patients
and third parties. Therefore, please observe the following instructions on your
product:
Never operate the device if it shows any obvious safety defects.
2.2 Danger for users, patients, and third parties
WARNING
Ensure that the device is always functional and ready for use!
Thepatientcouldsuocate.
Ensure that the device is always ready for use.
Place the device where it is easily accessible.
Make sure that the power cable/recharging unit incl. power cable is fully functional.
Always carry out a function check before using the device.
ATMOS recommends always having another suction device ready at hand.
Observe the notes on electromagnetic compatibility (EMC).
Only use the recommended original accessories and original spare parts.
Always use transparent suction hoses.
WARNING
Avoid misuse!
Risk of severe injury to your patient.
Employ the device only according to its intended use.
Never use the device for low-vacuum suction.
Too many suction operations may cause minor bleeding.
Always observe the applicable guidelines.
Using the device on children requires a lower vacuum setting. Observe the instruc-
tions issued by the attending physician.
The device must only be operated by persons who have been briefed on its medical
use.
Observe the accompanying operating instructions for all accessories.
Observe the notes on hygiene and cleaning.

Table of Contents