26
Operation
CAUTION
Risk of injury due to inappropriate material or untrained users!
Risk of injuries in the oral cavity and pharynx of the patient.
• Suction particularly carefully in the tracheal area.
• Thepatientmaysuerseriousinjuryiftoomuchvacuumisapplied.Turnthe
vacuum adjuster to set the vacuum to the required level. The vacuum gauge shows
the current vacuum level.
• Riskofsuocationwhensecretioncanistersystemisfull.
• Keep an eye on how full the secretion canister system is.
• Empty the secretion canister system as soon as it reaches half-full. When the
secretion canister system is full, you can no longer use the device for suction.
• Make sure that the suction hose is not kinked during suction. Otherwise, the suction
capacityappliedtothepatientwillbeinsucient.
• Insert the suction catheter as shown to you by the trained and instructed medical
professionals, and start the suction procedure.
• Control the suction procedure using the secondary air opening (
12
)onthengertip.
• ThehydrophobicDDSbacterialandvirallter/oversuctionstopsafelyprevents
liquid from getting into the pump. Nevertheless the rinsing canister must be
emptied when it is half full.
• Duetothedesignofthecanisterlid,theextractedsecretionowsalongthecanister
wall into the canister. This minimises foaming in the canister.
• Donotswitchothedeviceuntilyouhavenishedperformingsuction.
4.6 Hose rinsing
Dispose of the suction catheter and rinse the suction hose with clean water or disin-
fectant after every suction process. It is recommended to use an irrigation bottle in
which clean water can be carried.
4.7 DDS secretion canister system
ATTENTION
Important notes on safety for the DDS secretion canister system
• NeveroperatethedevicewithoutaDDSbacterialandvirallter.Alwayskeepat
leastonespareDDSbacterialandvirallteronhand.
• WeardisposablegloveswhenchangingtheDDSbacterialandvirallter.
• PriortoeachusecheckwhethertheDDSbacterialandvirallterisdryandclean.
ReplacetheDDSbacterialandvirallterwithanewDDSbacterialandvirallterifit
isdiscoloured,contaminatedoroversucked.TheDDSbacterialandviralltermust
not be dried and not reused.
• ReplacetheDDSbacterialandvirallterafterapatientchange.ReplacetheDDS
bacterialandvirallterafter14days,evenifthereisnopatientchange.