EasyManua.ls Logo

Atmos C 161 - Oversuction

Atmos C 161
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
Cleaning/disinfection
& If the device is still being used by the same patient, the surface should still be
cleaned at least once every week with a damp (not wet) cloth, and afterwards be
disinfected using one of the surface disinfectants listed in “5.5 Recommended
surface disinfectants” on page 39.
5.2 Oversuction
When is a suction device oversuctioned?
A suction device is oversuctioned if suction material has penetrated into the interior of
the device.
How can I see whether my suction device is oversuctioned?
Reduced suction capacity is usually a sign of potential oversuction. The ATMOS C 161
Aspirator, ATMOS C 261 Aspirator or ATMOS C 161 Battery also has a condensate
collector at the bottom of the device. Please remove the cover cap when performing a
visual inspection. The device is oversuctioned and contaminated if humidity or contami-
nation is visible in the condensate collector.
Measures
In case of oversuction or if there are any reservations regarding its hygienic condition,
thesuctiondevicemustbere-sterilisedbythemanufactureroracertiedATMOS
partner. A contaminated suction device constitutes a risk to both the patient and the
caregiver. As such, we recommend checking the condensate collector regularly.

Table of Contents

Related product manuals