Cleaning/disinfection
5.3 Cleaning instructions
For continued use on the same patient
Surface
For use on another
patient
If contaminated
After each suction
procedure
1x per day
1x per week
Every 2 weeks
Every 4 weeks
Exchange the DDS bacterial and
viral lter
X X X
Rinse the suction hose X X
Disinfect the suction hose X X
Exchange the suction hose X
Exchange the ngertip X X
Empty the secretion canister X X
Rinse the secretion canister X X
Disinfect the secretion canister X X
Rinse the canister lid components X X
Disinfect the canister lid components X X
Clean the device surface X X X
Wipe down the device surface with
disinfectant
X X X
Rinse the vacuum hose X X
Disinfect the vacuum hose X X
Disinfect the hose connector X X
5.4 Recommended instrument disinfectants
Disinfectant Ingredients (in 100g) Manufacturer
GIGASEPT FF
(Application concentrate)
Succindialdehyde
Dimethoxytetrahydrofurane
Corrosion inhibitors
non-ionic surfactants and
perfumes
11.0 g
3.0 g
Schülke & Mayr, Norderstedt
Sekusept PLUS
1
Glucoprotamine
Non-ionic tensides
Solvents, complexing agents
25.0 g Ecolab, Düsseldorf