For your safety
WARNING
Reduce the risk of infection for you and your patients!
Risk of deadly diseases being transmitted.
Always wear disposable gloves.
Never use components marked with
2
more than once.
Always use a suitable sterile catheter for suction. The suction hose must never
come into direct contact with the application site.
Only use sterile packaged parts if their packaging is undamaged.
Neveroperatethedevicewithoutabacterialandvirallter.Beforeeachuse,check
thatthebacterialandvirallterisdryandcleantoensurethatitoperatescorrectly.
WARNING
Protect yourself against an electric shock!
Damage from incorrect power supply.
Risk of burns, cardiac arrhythmias and even fatal injury.
ATMOS C 161 Aspirator and ATMOS C 261 Aspirator
- Prior to start-up, check whether the voltage and frequency data listed on the
device correspond to the values of the mains power supply.
ATMOS C 161 Battery
- Before switching on the device, make sure that the power supply grid is designed
for connecting a device that operates in the range from 100240 VDC at a grid
frequency of 50/60 Hz. Use only the included recharging unit for the device.
Do not operate the device in damp rooms, bathrooms or shower cubicles. Keep the
recharging unit, control panel and mains power connection point dry.
Do not use the device in areas where there could be a torrent of water.
Never immerse the device in water or other liquids.
Do not operate the device if it has been dropped. In this case, please clean the
device and send it in to ATMOS for repair.
Disconnect the device from the mains power supply prior to cleaning or disin-
fection.
Prior to each use, check for damage to the device and power cable/recharging unit
incl. power cable. Do not operate the device if you notice any damage. In this case,
please clean the device and send it in to ATMOS for repair.
Ensure that no liquid penetrates the device. If liquid gets into the device, operation
of the device must cease immediately. In this case, please clean the device and send
it in to ATMOS for repair.
The electrical suction devices are not sterilisable.
Only use the power cable/recharging unit incl. power cable in dry environments.
Always use the power cable/recharging unit incl. power cable in accordance with the
operating instructions.
Only use original accessories and original spare parts from ATMOS.
Observethespecicationsregardingperiodictestsinchapter“6.0Wartungund
Service” on page 42.
Assembly, new settings, alterations, extensions, and repairs may only be carried out
by authorised persons.