18 7821 User Instructions
Modes of Operation
11. Press the UNITS key to perform this test …
The display shows the weight on the scale at a resolution of 100,000
counts.
12. Press the UNITS key to stop the test …
HRESA is displayed.
13. Press the ZERO key …
HRESN is displayed. This stands for the high resolution test without AZT
enabled.
14. Press the UNITS key to perform this test …
The display shows the weight on the scale at a resolution of 100,000
counts.
15. Press UNITS key to stop the test …
HRESN is displayed.
16. Press the ZERO key …
LOC-G is displayed. This stands for local gravity or current gravity
setting.
17. Press the UNITS key …
The local or current gravity setting is displayed.
18. Press the ZERO key …
CAL-G is displayed. This stands for calibration gravity setting.
19. Press the UNITS key …
The calibration gravity setting is displayed.
20. When you are finished, press the ZERO key, which displays DONE, and press
the UNITS key, DIAG is displayed, or place Switch 1 back to the closed
position to return back to normal weighing mode.