EasyManua.ls Logo

Axon Axopatch 200B - Frequency Tuning (Whole Cell Config.)

Default Icon
153 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SPECIFICATIONS
107
Chapter 9
Reset Characteristics (Patch Mode only)
Total reset time
50 µs ± 10%. This includes:
integrator reset 10 µs
differentiator reset 30 µs
other overhead 10 µs
Time between resets (T
BR
)
For DC currents: T
BR
= 10/(I
DC
- I
BIAS
)
where I
DC
and I
BIAS
are in pA and T
BR
is in seconds.
I
BIAS
is typically 0.3 - 1.0 pA.
For transient currents: A reset will occur if the headstage must deliver more than 10 pC of
charge to the membrane. For example, a 60 mV step imposed on a 200 pF bilayer
membrane will cause a reset (12 pC of charge needed) whereas a 40 mV step will not (8 pC
of charge needed).
Reset transients in current waveform at Scaled Output (typical)
100 Hz bandwidth ± 0.25pA
1 kHz ± 0.5 pA
10 kHz ± 2 pA
Current Clamp
The current clamp mode has two speed settings: I-CLAMP NORMAL and I-CLAMP
FAST. I-CLAMP NORMAL is for use with electrode resistances greater than 1 M.
I-CLAMP FAST is for use with pipette resistances greater than 10 M. (The speed in I=0
mode is the same as in I-CLAMP NORMAL. In addition, TRACK mode is a slow clamp to
zero current.) Note that series resistance compensation remains active in current clamp
mode, allowing measurement of pipette resistance and (when R
s
is compensated) accurate

Table of Contents

Related product manuals