EasyManua.ls Logo

Axon Axopatch 200B - Panel Meter; Panel Meter Signals

Default Icon
153 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INSTRUMENT OPERATION
67
Chapter 6
In I-CLAMP mode, if the current is zero, 10 V
m
is ten times the actual membrane
potential. If current is flowing in the pipette, 10 V
m
includes the voltage drop
across the pipette. The actual membrane potential is given by the following
equation:
10 V
m
Actual Membrane Potential (mV) =
_______ _
I R
s
10
where I = pipette current in nA
R
s
= resistance in M set on SERIES RESISTANCE control
Panel Meter
The main panel meter displays one of five signals. Selection is made by a rotary knob. The
five signals are:
1) V
HOLD
/I
HOLD
: The value set on the HOLDING COMMAND potentiometer.
When V
HOLD
/I
HOLD
is selected, it will show the absolute value
of the HOLDING COMMAND even if the HOLDING
COMMAND control is OFF.
2) I: The pipette current. The reading is automatically scaled to
suit the headstage gain. The operation is autoranging; that is,
the decimal point and the units indicators automatically shift
so that large DC currents can be displayed.
3) I
RMS
: The rms current noise. The noise is measured in a 5 kHz
bandwidth using a Butterworth filter that is independent of the
front-panel lowpass Bessel filter. The front-panel gain does
not affect this reading.
4) V
m
: The membrane potential.
5) V
TRACK
: The output of the automatic offset nulling circuit.
6) TEMP: Headstage circuitry temperature in °C (range ± 20 °C).

Table of Contents

Related product manuals