EasyManuals Logo

Axon Axopatch 200B User Manual

Default Icon
153 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #131 background imageLoading...
Page #131 background image
SPECIFICATIONS
115
Chapter 9
Cell Capacitance
Telegraph Output: 0 to +10 V, proportional to setting 0-100 pF (for β = 1; 0-1000 pF for
β = 0.1 ) when WHOLE CELL CAP. switch is in the ON position.
0 to -10 V, when WHOLE CELL CAP. switch is in the OFF position.
Data Not Valid
Output goes High during a reset in PATCH mode or for the duration of a BLANK
ACTIVATE pulse in either PATCH or WHOLE CELL mode.
Panel Meter
3.5 digit meter displays TRACK potential (V
TRACK
) in mV, membrane potential (V
m
) in mV,
current noise (I
RMS
) in pA rms, membrane current (I) in pA or nA, HOLDING COMMAND
(V
HOLD
/I
HOLD
) in mV or nA, or capacitor-feedback circuitry temperature in degrees Celsius
(TEMP). Meter has autoranging feature when membrane current (I) or holding command
(V
HOLD
/I
HOLD
) is chosen.
Grounding
Signal ground is isolated from chassis and power ground. Signal ground is available on
rear panel.
Control Inputs
Above 3 V accepted as logic High. Below 2 V accepted as logic Low. Inputs protected to
±15 V.
Models
Unit is supplied with two model assemblies, the PATCH-1U and the MCB-1U.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Axon Axopatch 200B and is the answer not in the manual?

Axon Axopatch 200B Specifications

General IconGeneral
BrandAxon
ModelAxopatch 200B
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals