9-19
Chapter 9 Maintenance and Inspection
(2)  Attach the cover and inject silicone oil or distilled water with a syringe, etc., until it 
overflows
Fig. 9-29.  Adding silicone oil after attaching the cover
(3)  Remove the air remaining in the bolt holes in the cover with a clip.
(4)  Gently replace the rubber diaphragm while carefully checking that no air bubbles 
remain in back of it.
(5)  Add the ring and tighten the six bolts.
Tightening torque: 0.63 ± 0.1 N·m
Fig. 9-30.  Attaching the rubber diaphragm and tightening the bolts
  Handling Precautions
 
y
Exercise care that no air bubbles remain between the rubber diaphragm and the 
metallic diaphragm. If there are air bubbles, a measurement error may occur.
 
y
Make sure that the center of the rubber diaphragm is approximately 1 mm higher 
than its periphery.
Silicone oil  Bolt hole in cover (6 locations)
 Cover
Ring  Rubber diaphragm  Bolt