www.balluff.com 2
No. 873240 EN · I16; Subject to modification. Replaces K15.
english
Noise elimination
To avoid equipotential bonding - a current flow - through 
the cable shield, please note the following:
 – Use insulating bushes
 – Put the control cabinet and the system in which the 
transducer is located to the same ground potential.
Startup
  DANGER
Uncontrolled system movement
When starting up, if the position measuring system is 
part of a closed loop system whose parameters have not 
yet been set, the system may perform uncontrolled 
movements. This could result in personal injury and 
equipment damage.
 ► Persons must keep away from the system's 
hazardous zones.
 ► Startup must be performed only by trained technical 
personnel.
 ► Observe the safety instructions of the equipment or 
system manufacturer.
1.  Check connections for tightness and correct polarity. 
Replace damaged connections.
2.  Turn on the system.
3.  Check measured values and adjustable parameters 
regularly (especially after replacing the transducer or 
after repair by the manufacturer). Recalibrate the 
transducer, if necessary.
The calibration procedure is described in the 
comprehensive user's guide.
Operating notes
 – Check the function of the transducer and all associated 
components on a regular basis.
 – Take the position measuring system out of operation 
whenever there is a malfunction.
 – Secure the system against unauthorized use.
Installing the transducer (continued)
The transducer is electrically isolated from the machine 
with the supplied insulating bushes.
1.  Guide the transducer into the mounting clamps.
2.  Attach transducer to the base using mounting screws 
(tighten screws in the clamps with max.2Nm).
3.  Insert magnet (accessories).
Shielding and cable routing
Defined ground!
The transducer and the control cabinet must be 
at the same ground potential.
Shielding
To ensure electromagnetic compatibility (EMC), observe 
the following:
 – Connect the transducer and controller using a shielded 
cable. 
Shielding: Copper filament braided, at least 85% 
coverage.
 – Shield is internally connected to connector housing.
Magnetic fields
The position measuring system is a magnetostrictive system. 
It is important to maintain adequate distance between the 
transducer and strong, external magnetic fields.
Cable routing
Do not route the cable between the transducer, controller, 
and power supply near high voltage cables (inductive stray 
noise is possible).  
The cable must be routed tension-free.
Cable length
BTL7-A/G Max. 30 m1)
BTL7-C/E Max. 100 m1)
1) Prerequisite: Construction, shielding and routing preclude the effect of 
any external noise fields.
Electrical connection
Pin  BTL7-A501-… BTL7-G501-… BTL7-C501-… BTL7-E501-…
Pin assignment of S115 connector (view of 
connector pins of transducer)
1 0V (pin 3)
2 0V (pin 5)
3 10…0V1) 10…−10V1) 20…0mA1) 20…4mA1)
4 La (communication line)
5 0…10V1) −10…10V1) 0…20mA1) 4…20mA1)
1) Factory setting, can be freely configured with the 
PC software.
2) Reference potential for supply voltage and 
EMC-GND.
6 GND2)
7 10…30V
8 Lb (communication line)
BTL7-A/C/E/G501-M _ _ _ _ -P-S115
Micropulse Transducer in a Profile Housing