www.balluff.com 23english
b
a
b
8
  Adjusting (continued)
Adjust  
start value
(See page22)
Displayed values (example)
At 0 to 10 V At 4 to 20 mA
4. Adjust end value
 ► Bring magnet to end position.
9.89 V 19.13 mA
 ► Activate 
 for at least 2s.
> 2s  
 ⇒ Indication for “Adjust end value” is displayed.
10.00 V 20.00 mA
 ► Adjust end value
 ⇒ The end value can be changed using 
 and 
1)
. The gradient of the output is changed, 
but the null value remains unchanged (see 
page18).
9.89 V 19.13 mA
9.00 V 19.00 mA
 ► Exit calibration procedure: Activate 
 and 
 for  
no more than 2 s.
< 2s  
a
b
 ⇒ Indication for “Adjustment” is displayed.
2.00 V 6.00mA
 ⇒ Set position value is saved.
9.00 V 19.00 mA
Check values
The settings for the start value and end 
value have a mutual effect depending 
on the measuring position.
Repeat steps 3 and 4 until the desired 
values are exactly set.
5. Exit adjustment and deactivate buttons
 ► Simultaneously activate 
 and 
 for at least 6 s.
> 6s  
a
b
 ⇒ Output indicates error value.
10.50 V 3.60 mA
 ► Briefly activate 
 or 
 (< 1 s).
< 1s  
b
 ⇒ Buttons are deactivated.
 ⇒ Current position value is displayed.
9.00 V 19.00 mA
1) Briefly activate button: Current value is increased or 
decreased by approx. 1mV or 1 µA.
If a button is activated longer than 1 s, the step interval is 
increased.
BTL7-A/C/E/G5_ _-M_ _ _ _-J-DEXC-TA12
Magnetostrictive Linear Position Sensor – Rod Style
Type of protection “db” and “ta”
Flameproof enclosure