www.balluff.com 2
No. 880278 EN · A16; Subject to modification. Replaces edition 1206.
english
Only use the proper nut for the mounting thread. Firmly 
tighten the transducer with a maximum torque of 100Nm.
Radial cable outlet
During installation, the direction of the cable 
outlet is specified by the thread.
For horizontal assembly with nominal lengths > 500 mm, 
support the rod and tighten it at the end if necessary (only 
possible with a diameter of 10.2mm).
If installed in a hydraulic cylinder, the magnet should not 
make contact with the rod. Minimum bore diameter in the 
support piston:
Rod diameter Bore diameter
10.2mm At least 13mm
8mm At least 11mm
Startup
  DANGER
Uncontrolled system movement
When starting up, if the position measuring system is 
part of a closed loop system whose parameters have not 
yet been set, the system may perform uncontrolled 
movements. This could result in personal injury and 
equipment damage.
 ► Persons must keep away from the system's 
hazardous zones.
 ► Startup must be performed only by trained technical 
personnel.
 ► Observe the safety instructions of the equipment or 
system manufacturer.
1.  Check connections for tightness and correct polarity. 
Replace damaged connections.
2.  Turn on the system.
3.  Check measured values and adjustable parameters 
regularly (especially after replacing the transducer or 
after repair by the manufacturer). Recalibrate the 
transducer, if necessary.
The calibration procedure is described in the 
comprehensive user's guide.
Shielding and cable routing
Defined ground!
The transducer and the control cabinet must be 
at the same ground potential.
Shielding
To ensure electromagnetic compatibility (EMC), observe 
the following:
 – Connect the transducer and controller using a shielded 
cable. Shielding: Copper filament braided, at least 85% 
coverage.
 – Connector version: Shield is internally connected to 
connector housing.
 – Cable version: On the transducer side, the cable 
shielding is connected to the housing. Ground the 
cable shielding on the controller side (connect with the 
protective earth conductor).
Magnetic fields
The position measuring system is a magnetostrictive 
system. It is important to maintain adequate distance 
between the transducer cylinder and strong, external 
magnetic fields.
Cable routing
Do not route the cable between the transducer, controller, 
and power supply near high voltage cables (inductive stray 
noise is possible). The cable must be routed tension-free.
Cable length
BTL7-A/G Max. 30 m1)
BTL7-C/E Max. 100 m1)
1)  Prerequisite: Construction, shielding and routing preclude the effect of 
any external noise fields.
Installation
NOTICE!
Interference in function
Improper installation can compromise the function of the 
transducer and result in increased wear.
 ► The mounting surface of the transducer must make 
full contact with the supporting surface.
 ► The bore must be perfectly sealed (O-ring/flat seal).
Electrical connection
Pin Cable color -A510 -G510 -C500 -C570 -E500 -E570
Pin assignment of S32 (view of 
connector pins of transducer), 
8-pin M16 circular plug
1 YE yellow Not used 1) 0…20 mA 20…0 mA 4…20 mA 20…4 mA
2 GY gray 0 V
3 PK pink 10…0V 10…-10V 10…0V 3)
1) Unassigned leads can be connected to GND 
on the controller side but not to the shield.
2) Reference potential for supply voltage and 
EMC-GND.
3) The voltage outputs may exhibit a minimal 
offset to the harmonized current output.
4 RD red La (programming input)
5 GN green 0…10V -10…10V 0…10V 3)
6 BU blue GND 2)
7 BN brown 10…30V
8 WH white Lb (programming input)
BTL7-A/C/E/G5 __ -M ____ -H/W(8)-S32/___
Micropulse Transducer - Rod Style