3.3. Perform Material Investigation
Before investigation for unknown materials, the corresponding Operation Preset needs to
be selected.
Press the Set Up icon on the Home page, and then press Operation Preset. The list for
the existing Operation Preset will be displayed. Select the desired Operation Preset by
tapping on the name. The selected Preset will be marked by a check mark. Press on
View to view the setting or Back to return to the previous screen.
Note: Mixture MinHQI is not available for NanoRam Mini.
The Operator is only able to view the parameters. If a parameter is changed
unintentionally, the change will not be saved, thus the Preset remains intact.
Make sure the appropriate sampling adaptor is properly installed. Place sample correctly