1-5
1. Product Information
Important key commands, Continued
After pressing push-button switch: Signal lamps
High Voltage OFF
Testing Set on STANDBY
- The mode of operation changes from ACTIVATED to STANDBY.
- Primary circuits of high voltage supply are disconnected from
mains.
- Discharge device is connected to high voltage conducting parts.
red
green
illuminated
not illuminated
The Emergency-Stop pushbutton switch:
- is used for immediate shutdown in case of danger. (High
voltage will be switched off and discharge device is activated).
- is equipped with a key lock.
- Pressing the Emergency-Stop push-button switch and removal
of the key assures protection against unauthorized usage.
Emergency-Stop
push-button switch,
lockable
Use appropriate equipment:
- make sure that all parts that are not earthed (low resistance)
are no longer alive!
Before removing safety
precautions