5-4
5. Servicing
Depending on the frequency of usage, check periodically transient
protection resistor and discharge device using an Ohmmeter. The
total resistance R
S
must be between 60 and 80kW .
Check transient protection resistor
0
2
4
6
8
10
12
14
16
18
20
22
24
26
28
30
100mA
10mA
1mA100µA
10µA
1µA
F1
F2
PGK
T / min
U
DC
U
max
B
0
20
4
0
6
0
8
0
1
0
0
U [kV]
I
0
0
0
R
S
= 60 - 80 kW!
Measurements are only allowed with Testing Set switched off!
22 kW