Single needle – start the treatment
4/29/20
100
Once the treatment has been started, press the flashing arterial pressure control
to centralize the alarm limits around the current value.
The alarm limits for the venous pressure are automatically set once the treatment
has started to +350mmHg (high limit) and +150mmHg (low limit). Press the
venous pressure control if it is necessary to change the limits.
The stroke volume that can be achieved is determined by the size and position of
the venous pressure alarm window, and the potential expansion volume provided
by the expansion chamber.