Single needle - activate
After priming is completed, check the priming fluid level in the expansion chamber
is around the groove and return it to the upright position.
Press the blood button and open the blood menu. Press the single needle tab and
set a minimum stroke volume limit, and then activate single needle. The machine
is now ready for patient connection.