Troubleshooting – power failure
If there is a power failure, the machine has a
battery back-up which lasts approximately 30
minutes and keeps the screen and the blood
pump operational. During this time the dialysis
fluid is not heated.
If all the battery back-up charge is used up, or it
does not work for some reason, then the
machine will shut down. All the settings and the
actual values will be kept.
When the power returns, press the On/Off button
to start the machine. The machine will perform a
recovery and the treatment will continue from
where it stopped. However, all the treatment
parameters must be checked when starting after
a recovery.
4/29/20
87
Battery connect indicator is lit green if the
battery back-up has been installed
Battery charge indicator is lit yellow when
the mains cable is connected to the mains
supply and the main switch is switched on