The blood can be returned manually in the event of a power failure when no
battery back-up is available.
Connect arterial blood line to the rinse-back solution
Open the blood pump cover
Manually turn the blood pump anti-clockwise
Troubleshooting manual rinse back
4/29/20
88
When returning the blood manually, the operator is responsible for
visually observing all safety parameters