• With a gloved hand,
place the BD Microtainer
®
Quikheel
™
Lancet against
the site with the logo
facing you.
• Place the blade slot area
securely against the heel.
• Place the incision at a 90º
angle to the length of the
foot.
• Select a site identified in one
of the colored areas shown
in the illustration to the left.
• Cleanse site thoroughly
with alcohol (70/30 by
volume, 70%).
• Allow skin to air dry.
BD Microtainer
®
Quikheel
™
Lancet
NCCLS LA4-A3 and H4-A4 Recommended Guidelines for
performing a heel stick:
1. Warming the heel increases arterial blood flow to the site up
to seven-fold without burning the skin.
2. Temperature of warming device should be no higher than
42ºC (107.6°F).
3. Warm for 3 minutes.
1
2
• Firmly and completely depress the trigger with
your index finger.
• After triggering the lancet, remove the lancet and
discard it into a biohazard sharps container.