BD Phoenix™ M50 System Quick Reference Guide
Panel Login
1 Press the Panel Login tab and select Clinical.
2 To receive special notification when results are obtained, select Critical.
3 In the Accession Number field, type in/scan an accession number.
4 In the Sequence Number field, type in/scan the panel’s sequence number.
5 The Isolate Number field defaults to isolate number 1. Type in the isolate number. Valid isolate numbers are 1 to 20.
6 If only the ID or AST portion of a combination panel is being used, disable the part of the panel not being used.
7 If ID is disabled, or if an AST only or BD Phoenix™ EMERGE™ panel is not being used, the Organism ID field appears.
8 For Yeast panels, select the media type in the media field.
9 Save the information.
10 Insert the panel in the instrument (refer to Section 5.2.3 of the BD Phoenix™ M50 System User’s Manual).
Panel Login