BeaconMedaes Gemini - Tools and Service Materials; Cleaning and Lubrication; Inspection and Testing

BeaconMedaes Gemini
12 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4
MAN01-049
Part No. 847706-00 Rev. C00
Description Manufacturer
Lubricant Krytox® GPL206
BeaconMedæs
P/N GRS01-007
No. 2 Phillips screwdriver Local source
No. 2 Flat-blade screwdriver Local source
Oxygen compatible leak Sherlock or equivalent
detector solution
Cleaning and Lubrication
BeaconMedæs outlets are cleaned for oxygen or the
indicated service. Exposed surfaces of outlets may be
cleaned with a mild detergent solution or wiped with
a disinfectant used for patient room surfaces that is
compatible with acetal and nylon plastic, anodized
aluminum, and die cast zinc.
Lubricate elastomer seals sparingly with Kryto
GPL206. Refer to Tools and Service Materials.
DO NOT USE OIL.
Inspection and Testing
Medical gas outlets in high-use areas should be
inspected at least every six months and more frequently
if outlets are used daily. All outlets should be inspected
at least once a year.
Follow National Fire Protection Association
Standard 99, 99C “Gas and Vacuum Systems,
and Canadian Standards Association, Standard Z
305.1 “Nonfl ammable Medical Gas Piping Systems”
requirements and methods for periodic testing and
maintenance of medical gas and vacuum systems.