Troubleshooting Guide
Symptom Possible Cause Corrective Action
Quick-Connect or DISS Valve
Leaks...
• without quick-connect
adapter in place
Faulty primary seal Replace primary seal
Faulty rubber bushing Replace rubber bushing
with quick-connect adapter in
place
Faulty primary seal Replace primary seal
Faulty rubber bushing Replace rubber bushing
Faulty quick-connect stem Replace quick-connect stem
without DISS nut and nipple
in place
Faulty primary seal Replace primary bushing
Faulty rubber bushing Replace rubber bushing
with DISS nut and nipple in
place
Faulty DISS nipple
Faulty O-Ring seal
Replace DISS nipple
Replace O-Ring seal
Adapter will not engage and/or
release
• Damaged adapter • Replace adapter
Wrong adapter Check adapter/outlet gas
indexing and labels
Damaged frontbody assembly Replace frontbody assembly
Flow rate does not meet
specifi cation
Piping system pressure/
vacuum incorrectly set
Reset system pressure/
vacuum
Restriction in piping system Check system and/or outlet
System demand greater than
supply source or system
capacity
Check supply source and
system sizing
Outlet pressure or vacuum does
not meet specifi cation
System pressure/vacuum not
set correctly
• Reset
System demand exceeds
supply source capability or
system capacity
Check supply source and
system sizing
Restriction in piping system or
outlet
Check and remove