Beckman Coulter Airfuge - User Manual

Beckman Coulter Airfuge
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
AF-IM-13AC
June 2020
Instruction For Use
Airfuge
Air Driven Ultracentrifuge
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea, CA 92821 U.S.A.

Table of Contents

Other manuals for Beckman Coulter Airfuge

Questions and Answers

  • B
    Benjamin HuberSep 9, 2025
    How to fix oil visibly collecting in Beckman Coulter Airfuge Laboratory Equipment?
    • C
      Christopher MckaySep 9, 2025
      If you notice oil visibly collecting in your Beckman Coulter Laboratory Equipment, the filter element is likely saturated and needs to be replaced.
  • A
    audreytuckerAug 23, 2025
    Why does Beckman Coulter Laboratory Equipment rotor run erratically?
    • P
      pgillAug 23, 2025
      If your Beckman Coulter Laboratory Equipment rotor runs erratically or makes unusual noises, the stator pad may be damaged and needs replacement. Other potential causes include dirty air jets (clean the stator assembly), a missing or damaged rotor cap (check the rotor cap), or an improperly loaded rotor (check placement of tubes or liner in rotor).
  • J
    Jeremy GrahamAug 25, 2025
    What to do if Beckman Coulter Laboratory Equipment rotor fails to stop?
    • M
      Michael AguilarAug 25, 2025
      If the rotor fails to stop in your Beckman Coulter Laboratory Equipment, the brake pin shaft may be oily, so clean the pin and shaft. Also, the levitation air flow might be too high, so adjust the levitation air. Finally, the brake spring tension could be too low, so adjust the brake spring tension.
  • P
    Paul JohnsAug 27, 2025
    Why does Beckman Coulter Airfuge rotor stop too fast?
    • S
      Sheryl SandersAug 27, 2025
      If the rotor stops too fast on your Beckman Coulter Laboratory Equipment, adjust the levitation air, as the levitation air flow may be too low. Also, check the time-delay setting because the time-delay period may be too short. Other possible causes include a worn or damaged stator pad (replace the stator pad), an oily brake pin shaft (clean the stator assembly and shaft), or a brake malfunction (adjust brake spring tension and levitation air).
  • A
    Angela DavidsonAug 28, 2025
    Why is oil visibly collecting in Beckman Coulter Airfuge Laboratory Equipment?
    • N
      nthompsonAug 28, 2025
      If you see oil visibly collecting in your Beckman Coulter Laboratory Equipment, the filter element is likely saturated and needs to be replaced.
  • M
    Marvin CooperAug 29, 2025
    Why pressure at Beckman Coulter Airfuge will no longer read 30 psig?
    • M
      Mark WeberAug 29, 2025
      If the pressure at your Beckman Coulter Laboratory Equipment no longer reads 30 psig, the filter element is likely saturated and needs to be replaced.
  • C
    Casey KellerSep 18, 2025
    What to do if brake pin not actuating in Beckman Coulter Laboratory Equipment?
    • T
      Tamara AdamsSep 18, 2025
      If the brake pin isn't actuating on your Beckman Coulter Laboratory Equipment, the brake spring tension may be too low, requiring adjustment. Also, the stator shaft might be oily, so clean the stator shaft and pin. If the brake pin is damaged, it will need to be replaced.
  • M
    Michael DavisSep 19, 2025
    Why does the brake pin fall flat on the stator surface of my Beckman Coulter Airfuge Laboratory Equipment?
    • B
      bphillipsSep 22, 2025
      If the brake pin is falling flat on the stator surface in your Beckman Coulter Laboratory Equipment, it could be because the brake spring wire isn't centered, needing adjustment. Alternatively, the brake spring tension might be too low, which also requires adjustment.
  • D
    David PerkinsSep 7, 2025
    Why does my Beckman Coulter Airfuge Laboratory Equipment rotor run erratically or make noise?
    • L
      Lauren FloresSep 8, 2025
      Erratic rotor behavior or unusual noises in your Beckman Coulter Laboratory Equipment can stem from several issues. A damaged stator pad could be the culprit, requiring replacement. Dirty air jets can also cause this problem, so cleaning the stator assembly is advised. Check for a missing or damaged rotor cap. Also, verify that the tubes or liner are correctly placed in the rotor.
  • C
    Cathy TaylorSep 4, 2025
    What to do if run light fails to come on in Beckman Coulter Airfuge?
    • N
      Nicole WilliamsSep 4, 2025
      If the run light fails to come on in your Beckman Coulter Laboratory Equipment, replace the lamp if it's burned out, or reset the circuit breaker if it has tripped. For a faulty time delay, call Beckman Coulter Field Service.

Summary

Safety Notice

Electrical Safety

Details electrical safety precautions to prevent injury and damage.

Mechanical Safety

Covers safe operational practices to prevent mechanical injury or damage.

Chemical and Biological Safety

Addresses safe handling of chemical and biological materials during operation.

Description

Controls and Indicators

Identifies and describes the function of the instrument's controls and indicators.

Installation

Space and Location Requirements

Details the necessary space and environmental conditions for installation.

Electrical Requirements

Specifies the electrical power requirements for the instrument.

Operation

Starting the Run

Explains the steps involved in starting an ultracentrifuge run.

Ending the Run

Details the procedure for safely stopping and finishing an ultracentrifuge run.

Troubleshooting

Troubleshooting Guide

Lists common malfunctions, their causes, and recommended corrective actions.

Maintenance

Beckman Coulter Airfuge Specifications