– 22 –
behr S 5
249101
Ending the manual distillation
Press the operating control knob to end the
distillation.
Start titration on your titrator. See user‘s manual of your
titrator.
Proceed ==>> to the next menu.
Select DrainSample. Hold the operating control knob
depressed until there is no more liquid in the sample
vessel.
Wait until the titrator has finished the analysis.
Select DrainTitrCell. Hold the operating control knob
depressed until there is no more liquid in the titration
vessel.
Caution! Sample vessel is hot!
After distillation do not touch the sample
vessel by hand. Use the crucible tongs or a
suitable kind of protection.
Depress the quick-release fastener and remove the
sample vessel.
Push the PTFE steam delivery tube into its retaining
notch.
By repeatedly pressing <<== you will return to the main
menu.
Distilling