What to do if ...
How to Display Current Operating Data
Dist.time
123
P0.280 f 6.8 I 69
While you are running a distillation, you can have a status
bar displayed that will show you some operating data:
on the left: the steam generator pressure in bar
in the midst: the signal frequency of the level sensor,
in kHz (low frequency = high level)
on the right: the heating current, in arbitrary units.
In order to display the status bar, just turn the operating
control knob, no matter if left or right.
If you turn the operating control knob once mire, the sta-
tus bar will vanish again.
This will work during a distillation only. While you are in
a menu, turning the operating control knob serves for
changing from a menu item to another or for modifying
an entry.
Error messages
The following error messages inform you that you must
take some corrective action in order to continue opera-
tion of the steam distillation apparatus:
Protection door open
You have attempted to execute a distillation program or
to generate steam by means of a MANUAL ENTRY, but
the protective door is open; or you have opened the pro-
tective door during a distillation.
Close the protective door, and press the control knob.
Distillation will continue.
Sample Vessel Not Inserted
You have attempted to execute a distillation program
or to generate steam by means of a MANUAL ENTRY,
without having first inserted a sample vessel into the
apparatus.
Using the MAIN SWITCH, turn the steam distillation
apparatus off.
Insert the steam delivery tube into an empty sample
vessel. Press the quick release fastener down and
insert the sample vessel.
Close the protective door and turn the apparatus on
using the MAIN SWITCH.
Insufficient Cooling-Water Pressure
The cooling water tap is closed or not opened sufficiently.
Using the MAIN SWITCH, turn the steam distillation appa-
ratus off.
Open the cooling water tap.
Turn the apparatus on again using the MAIN SWITCH .