EasyManua.ls Logo

Benning ST 725 - Page 38

Benning ST 725
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
07/ 2018
BENNING ST 725
30
BENNING ST 725 in mains operating mode:
- The BENNING ST 725 interrupts the testing procedure after the R
ISO
(insulating resistance)
measurement and requests the user to switch the mains voltage of 230 V to the test socket
1
by showing a flashing "I
LEAK
" symbol. Make sure that the test sample is protected and
press the -key
4
to measure the protective conductor current by means of the differential
current measurement method.
- The protective conductor current measurement (differential current measurement method)
only starts as soon as the mains voltage is correct applied (see chapter 8.3).
step 1 of 2:
- After a measuring time of 5 seconds or by pressing the key
4
, mains polarity will be
reversed and the protective conductor current will be measured with reversed mains voltage
("L/N" - "N/L"). The highest measured value of both measurements will be displayed. (step 2
of 2)
step 2 of 2:
- If the protective conductor current is lower than the admissible limiting value, a will be
shown next to the I
LEAK
symbol.
- The overall test is considered to be passed, if "PASS" is shown on the display.
As an alternative:
BENNING ST 725 in battery operating mode (without mains supply):
- Similarly, a will be shown next to the I
EA
symbol, if the protective conductor current I
EA
(al-
ternative leakage current measurement method) is lower than the admissible limiting value.
- The test is considered to be passed, if "PASS" is shown on the display.
Seegure4: TestingofdevicesofprotectionclassI(deviceswithprotectiveconductorand
accessible conductive parts which are connected to the protective conductor)
Note on measuring the protective conductor resistance:
- Alternatively, the measurement of the protective conductor resistance R
PE
can be carried out
as permanent measurement (max. 2 x 90 seconds). For this purpose press the key
2
for
approx. > 5 seconds until the symbol appears on the display. Check the connecting line of
the test object by bending it over the entire length in order to detect weak points or a break of
the protective conductor. The BENNING ST 725 continuously records the current measured
value on the display and stores the maximum value in the memory. By pressing the key
2
again, the measurement is carried out with reversed polarity. Press the key
2
again to
indicate the maximum value of R
PE
on the display and to continue the testing procedure as
described in section 9.1.
Note on measuring the protective conductor current in mains operating mode:
- Alternatively, the measurement of the protective conductor current I
LEAK
can be carried
out as permanent measurement (max. 2 x 5 minutes). Press the key
4
for approx. > 5
seconds to start permanent measurement. After 5 minutes, the polarity of the mains voltage
will be reversed automatically ("L/N" - "N/L"). By pressing the key
4
earlier, the mains
voltage polarity reversal can be activated manually and by pressing the key
4
again, the

Table of Contents

Other manuals for Benning ST 725

Related product manuals