Evaluation unit
LB 452
Operating Manual
47344BA2
Rev. No.: 04, 03/2015
Embedded Software as of vers. 1.3.0 (CU) and 1.3.0 (MU)

Table of Contents

Questions and Answers

  • J
    James LeeSep 18, 2025
    What to do if there is Quartz synchronization error on Berthold castxpert LB 452 Measuring Instruments?
    • T
      Terri DunlapSep 18, 2025
      If you are experiencing issues with quartz synchronization on your Berthold Measuring Instruments, contact service if the error persists.
  • G
    Grant WilsonAug 25, 2025
    What to do if the fill level display deviates in Berthold Measuring Instruments?
    • K
      Kelly MooreAug 25, 2025
      If the fill level display on your Berthold Measuring Instruments is deviating, check the detector and calibration values. Also check the source age and irradiation level, replace the detector if needed.
  • C
    Christopher StokesAug 25, 2025
    Why is the count rate too low on Berthold castxpert LB 452 Measuring Instruments?
    • M
      Matthew SmithAug 26, 2025
      If the count rate is too low on your Berthold Measuring Instruments, first check if the shielding is correctly opened. Ensure the lock is in the OPEN position. Verify and optimize the alignment to correctly focus the radiation on the detector. Offset any objects in the beam path. If the source is old, consider replacing it.
  • C
    christina03Aug 27, 2025
    What to do if no data is being received in the control center via Ethernet from Berthold Measuring Instruments?
    • Y
      Yolanda SmithAug 27, 2025
      If you are not receiving data in the control center via Ethernet with your Berthold Measuring Instruments, activate the output of measurement data over the network in the Data Logger menu under Network settings.
  • M
    Matthew DavisSep 19, 2025
    How to resolve software exception on Berthold castxpert LB 452?
    • S
      Steve RitterSep 22, 2025
      If you are experiencing a software exception with your Berthold Measuring Instruments, carry out a software update. If the error persists, contact the service.
  • R
    Rebecca HeathSep 23, 2025
    What to do if Decay compensation fails on Berthold Measuring Instruments?
    • L
      Lauren CabreraSep 23, 2025
      If decay compensation is failing on your Berthold Measuring Instruments, contact service.
  • P
    Paul BaileySep 9, 2025
    What to do if Calibration is not monotonic in Berthold castxpert LB 452?
    • C
      Christina OwensSep 9, 2025
      If the calibration is not monotonic on your Berthold Measuring Instruments, check the calibration table.
  • M
    Matthew CruzSep 12, 2025
    How to fix count rate stability out of bounds on Berthold castxpert LB 452 Measuring Instruments?
    • A
      Amanda RandallSep 12, 2025
      If the count rate stability is out of bounds on your Berthold Measuring Instruments, ensure that no other factors influence the count rate during the read-in (e.g., disruptions in the beam path). Carry out the adjustment again (twice, if needed). The error is automatically cleared.
  • J
    Justin ClarkSep 3, 2025
    What to do if WD off occurs in Berthold Measuring Instruments?
    • G
      Gloria BrownSep 3, 2025
      If the watchdog is disabled on your Berthold Measuring Instruments, remove the debug jumper.
  • J
    Jason WuSep 1, 2025
    How to resolve device data inconsistency found on Berthold castxpert LB 452?
    • I
      iwilliamsSep 1, 2025
      If your Berthold Measuring Instruments are showing an issue with device data, carry out a factory reset.

Summary

Safety

Proper Use of the Evaluation Unit

Details on adhering to instructions and ensuring safe and reliable operation of the EVU.

Improper Use and Associated Dangers

Highlights risks of improper use, including electric shock and personal injury.

Qualification of Personnel for Operation

Specifies the minimum requirements and groups of personnel for safe handling and maintenance.

Product Description

Radiometric Measurement System Principles

Explains the functional principle and components of the radiometric measurement setup.

Start-Up Procedures

Installation of the Evaluation Unit (EVU)

Step-by-step guide for installing the EVU, including safety precautions.

Configuring the Current Output

Instructions for setting up current outputs on basic and extension modules.

Operation of the Software

Overview of the Standard Display

Description of the main display screen and its elements during operation.

The System Menu

Channel Overview Menu Functions

Troubleshooting

Error Search and Common Problems

Guide to identifying and resolving common issues with the EVU.

Error Codes of the Evaluation Unit

Detailed list of system error codes, their descriptions, and corrections.

Berthold castxpert LB 452 Specifications

General IconGeneral
Detector TypeSilicon Drift Detector (SDD)
InterfaceUSB, Ethernet
Power Supply100-240 V AC, 50/60 Hz
Measurement PrincipleX-ray fluorescence (XRF)
Measurement Time10 to 300 seconds (depending on application)
X-ray Tube Voltageup to 50 kV