15 TROUBLESHOOTING
When starting the AlphaGUARD, several routines for system check are carried out.
If an error is detected, the associated error code is displayed according to
following table:
Severe Error, device no longer operational
0x00000001
Fail to initialize system
oscillator (not used at
the moment)
Internal Hardware Error ! , contact
your local Bertin
dealer, press <enter>
to turn off device
0x00000002
Clock hardware not
available
Internal Hardware Error ! , contact
your local Bertin
dealer, press <enter>
to turn off device
0x00000004
EEPROM chip not
writeable (ID Read
returned only 0x00 or
0xFF)
Internal Hardware Error ! , contact
your local Bertin
dealer, press <enter>
to turn off device
0x00000008
FLASH chip not
readable (ID Read
returned only 0x00 or
0xFF)
Internal Hardware Error ! , contact
your local Bertin
dealer, press <enter>
to turn off device
0x00000010
ADC converter init
failed
Internal Hardware Error ! , contact
your local Bertin
dealer, press <enter>
to turn off device
Severe Error, device still operational
0x00000100
Invalid parameter
values found, firmware
defaults loaded
System parameters corrupted, default
values from firmware loaded, contact
your local Bertin
dealer, press enter to
continue
0x00000200
Invalid parameter
values found, backup
defaults loaded
System parameters corrupted, default
backup values loaded, contact your
local Bertin dealer, press enter to
continue
0x00000400
No valid parameter
Backup available
No valid system backup available,
contact your local Bertin dealer, press
enter to continue
0x00000800
Tamper sensor init
failed
Disabled tamper measurement
0x00001000
pressure sensor init
failed
Disabled pressure measurement