77 WWW.BERTIN-INSTRUMENTS.COM
Basic functions AlphaGUARD DF2000
Ionization chamber, HV | 750 VDC
3D-alpha spectroscopy a n d current mode
0,62 liter (38 cubic inches)
0,56 liter (34 cubic inches)
Type of Radon progeny filter
(detector entry window)
Fine dust
filter (retention coefficient >99,9 %)
Detector filling mechanism
Design-optimized for fast passive diffusion
Transient response function
(time delay)
Signal > 30 % after 10 min
Signal > 70 % after 20 min
Signal > 90 % after 30 min
Detector signal acquisition
Fast digital signal sampling network, using
three separate ADC channels
Spectral signal extraction
DSP (Digital Signal Processing), on-line-cross
correlation algorithms
Detector efficiency in Radon
measurement mode
1 CPM at 20 Bq/m³ (0,55 pCi/l)
Detector efficiency in Rn/Tn
discrimination mode
- for radon
- for thoron at 1l/min flow rate
- for thoron at 2l/min flow rate
1 CPM at 60 Bq/m³ (1,6 pCi/l)
1 CPM at 200 Bq/m³ (5,5 pCi/l)
1 CPM at 140 Bq/m³ (3,8 pCi/l)
Background due to internal
detector contamination
Measurement range Radon
(Rn-222)
2 Bq/m³ … 2 000 000 Bq/m³
(0.05 pCi/l … 54 000 pCi/l)
Measurement range Thoron
(Rn-220)
2 Bq/m³ … 2 000 000 Bq/m³
(0.05 pCi/l … 54 000 pCi/l)
> 10.000.000 Bq/m³ (> 300 000 pCi/l) verified
Instrument calibration error
(Rn-222)
r
3 % (plus uncertainty of the primary
standard)