60 WWW.BERTIN-INSTRUMENTS.COM
AquaKIT
In combination with AlphaGUARD DF2000 the AquaKIT
radon concentration in water samples directly and precisely. The glass vessels of
the AquaKIT measuring equipment grant a hermetically sealed enclosure of the
radon, expelled from the water samples as well as a fast change of samples -
which prevents incorrect measurements as a consequence of leakages.
With the AquaKIT set-up, the AlphaGUARD DF2000 user can determine radon
directly and radium indirectly in water samples. In a closed gas cycle radon is
expelled from the water sample by the internal pump of AlphaGUARD. With the
AlphaGUARD DF2000, whose ionization chamber is also part of the gas cycle, the
radon concentration in the system is determined and stored in its memory (1
- resp.
10- min. cycle) in form of a concentration time series.
Figure 52: Water sample measurement with AquaKIT
For details regarding the simultaneous measurement of radon decay products
please see the AlphaPM Instruction Manual.
For details regarding the simultaneous measurement of radon decay
products please see the AquaKIT Instruction Manual.