79 WWW.BERTIN-INSTRUMENTS.COM
Multiparameter Features AlphaGUARD D50, D2000,
DF2000
Atmospheric air pressure
- Type of sensor
- Measurement range
- Resolution displayed on Screen
- Resolution under DataVIEW or
DataEXPERT
- Initial calibration uncertainty
Piezo-resistive semiconductor sensor
700 mbar ... 1100 mbar
0,1 mbar
0,1 mbar
+/- 3 mbar
Ambient temperature (sensor in
ionization chamber)
- Type of sensor
-
Measurement range
- Resolution displayed on Screen
-
Resolution under DataVIEW or
DataEXPERT
- Initial calibration uncertainty
Band gab semiconductor sensor
- 20 °C ... + 70 °C (- 4 ... + 158 °F)
0,1 °C (0,1 °F)
0,1 °C (0,1 °F)
+/- 1,5 °C (+/-2,5 °F)
Relative air humidity (sensor in
ionization chamber)
- Type of sensor
- Measurement range
- Resolution displayed on Screen
- Resolution under DataVIEW or
DataEXPERT
- Initial calibration uncertainty
Capacitive semiconductor sensor
0 % rH ... 99 % rH
0,1 % rH
0,1 % Rh
+/- 3 % rH
External sensor channel 1 & 2
Measurement range
Resolution under DataVIEW or
DataEXPERT
Signal sampling rate
Total signal error
Input impedance
Input connector type
0 VDC ... 2,5 VDC
0,00061 VDC
30 per minute
+/- 0,01 VDC plus +/- 3 %
10 kOhm
HIROSE HR10A-10R-10PB