EasyManua.ls Logo

Bertin Technologies ALPHAGUARD DF2000 - 12.4 Multiparameter Features AlphaGUARD D50, D2000, DF2000

Default Icon
113 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
79 WWW.BERTIN-INSTRUMENTS.COM
USER MANUAL
Multiparameter Features AlphaGUARD D50, D2000,
DF2000
Atmospheric air pressure
- Type of sensor
- Measurement range
- Resolution displayed on Screen
- Resolution under DataVIEW or
DataEXPERT
- Initial calibration uncertainty
Piezo-resistive semiconductor sensor
700 mbar ... 1100 mbar
0,1 mbar
0,1 mbar
+/- 3 mbar
Ambient temperature (sensor in
ionization chamber)
- Type of sensor
-
Measurement range
- Resolution displayed on Screen
-
Resolution under DataVIEW or
DataEXPERT
- Initial calibration uncertainty
Band gab semiconductor sensor
- 20 °C ... + 70 °C (- 4 ... + 158 °F)
0,1 °C (0,1 °F)
0,1 °C (0,1 °F)
+/- 1,5 °C (+/-2,5 °F)
Relative air humidity (sensor in
ionization chamber)
- Type of sensor
- Measurement range
- Resolution displayed on Screen
- Resolution under DataVIEW or
DataEXPERT
- Initial calibration uncertainty
Capacitive semiconductor sensor
0 % rH ... 99 % rH
0,1 % rH
0,1 % Rh
+/- 3 % rH
External sensor channel 1 & 2
-
Measurement range
-
Resolution under DataVIEW or
DataEXPERT
-
Signal sampling rate
-
Total signal error
-
Input impedance
-
Input connector type
0 VDC ... 2,5 VDC
0,00061 VDC
30 per minute
+/- 0,01 VDC plus +/- 3 %
10 kOhm
HIROSE HR10A-10R-10PB