23 WWW.BERTIN-INSTRUMENTS.COM
USER MANUAL
Acquisition
By means of five acquisition windows all important information on the
configuration as well as current measurement is available. Via pushing the Scroll
up or Scroll down button .
7.1.1 Acquisition 1/5
Displays the current measurement values of the radon concentration in Bq/m³ (or
pCi/l),
the
temperature in °C (or °F), the relative air humidity in %rH, the air pressure
in mbar or hpa (or inHg) and the gamma ambient dose rate in nSv/h (if equipped
with dose rate module). When operating in Radon/Thoron discrimination mode,
the thoron value appears on
Acquisition 1/5 instead of the gamma ambient dose
rate (DF2000). The radon and the thoron values are followed by the appropriate
statistical error bars.
Figure 10: Acquisition window 1/5
In order to record and
display
gamma dose rate measurement values it is required
to equip the AlphaGUARD with an additional module consisting of a Geiger-Müller
tube and associated electronics. The dose rate (gamma) as ambient dose
equivalent rate H
*10 is displayed in the unit nSv/h. The measuring range of the
additional gamma dose rate detector is 20 nSv/h - 10 mSv/h.
The display value for the gamma dose rate is updated along with each
new measuring cycle.