EasyManua.ls Logo

Bertin Technologies ALPHAGUARD DF2000 - Page 76

Default Icon
113 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
76 WWW.BERTIN-INSTRUMENTS.COM
USER MANUAL
Background signal due to
internal detector
contamination
< 1 Bq/m³ (0.03 pCi/l)
Measurement range Radon
(Rn-222)
2 Bq/m³ … 2 000 000 Bq/m³
(<0.05 pCi/l … 54 000 pCi/l)
Resolution on LCD display
1 Bq/m³ (0.01 pCi/l)
Fold back protection
> 10.000.000 Bq/m³ (> 300 000 pCi/l) verified
System linearity error
< 3% within total range
Instrument calibration error
(Rn-222)
r
3 % (plus uncertainty of the primary
standard)
Measurement modes and
cycles
10 min, 60 min (Diffusion)
Data capacity (non-volatile)
Up to 60 000 measurement points
~ 400 days at 10 min measuring cycle
~ 2 500 days at 60 min measuring cycle
Resolution of graphic display
160 x 104 pixels
Operation from internal
battery
> 10 days
Weight (incl. Internal battery)
6,2 kg (13,7 lbs)
Dimensions without handle
(L x W x H)
Dimensions with handle
(L x W x H)
282 mm x 340 mm x 123 mm
329 mm x 355 mm x 123 mm
System operating range
-
Temperature
-
Atmospheric pressure
- Humidity
-10 °C … +50 °C (+14 … +122 °F)
700 mbar … 1100 mbar
0 % rH … 95 % rH (non-condensing)
Power consumption external
power supply
100 – 240 V (400 mA)
50 – 60 Hz

Related product manuals