25 WWW.BERTIN-INSTRUMENTS.COM
USER MANUAL
7.1.3 Acquisition 3/5
This window allows the simultaneous chart of two data values. The last 143
measurement values of each parameter can be displayed simultaneously. The
time base depends on the chosen measuring cycle: With a measuring cycle of
1 min, the values of the last 143 min, with a measuring cycle of 10 min, the last 24
hours and with a measuring cycle of 60 min, the last 6 days can be displayed.
Figure 12: Acquisition window 3/5
Whenever
the data series exceeds the limit of 143 values, which can be displayed
in one window, the scroll mode can be activated by pushing shortly the Enter
button . The Scroll up and Scroll down buttons can be used to navigate through
the complete time period of the measurement experiment. The scroll mode can
be left by pushing the Enter button again.
13.08.12 12:23
12.08.12 12:23
+0
+20000
+0
+7000
5 min
Scroll mode 1/2
3630 Bq/m³Rn
11405 Bq/m³Tn
1- -1
2- -2
Figure 13: Scroll mode