40 WWW.BERTIN-INSTRUMENTS.COM
Acquisition 1/5
11.08.16 12:27:47 4min
Rn 3630± 687 Bq/m³
Tn 11405±3490 Bq/m³
T
A
+18.6 °C
rH 63 %rH
P
A
1004 mbar
!
Figure 40: Alarm indication by exclamation mark (!) on the top right corner
7.2.9.6 Set units
The units of the radon concentration, the temperature and the pressure can be
changed by using the Scroll up or the Scroll down button. The next field can be
selected by pushing the Enter button shortly. The selection has to be
acknowledged by pushing the Enter button for 2 sec.
The following units are selectable:
x
Radon concentration: Bq/m³, Ci/l
x Temperature: °C, °F
x Pressure: mBar, hPa, inHg
Set units
11.08.16 12:27:47 4min
°C
Temperature:
Pressure:
mbar
Radon konzentration:
Bq/m³
Figure 41: Changing the units for radon concentration, temperatur
The alarm is disabled when reducing the alarm threshold to values < 2 Bq/m³.