43 WWW.BERTIN-INSTRUMENTS.COM
”communication module” of the DataVIEW or DataEXPERT software recognizes
the set baud rate of the AlphaGUARD automatically.
The transfer of a several days lasting measuring series out of the value storage of
the AlphaGUARD into the working memory of a PC lasts with 115 200 Bd some
seconds, max. a few minutes depending on amount, operation mode, etc. With
appreciably lower baud rates this process can take several
minutes more. The set
baud rate can be changed by selecting the required rate and pressing the Enter-
key for acknowledgement.
Set RS232baudrate
11.08.16 12:27:47 4min
1200 Bd
300 Bd
4800 Bd
115200 Bd
38400 Bd
Use ↑↓ to change and
press <┘key to accept !
9600 Bd
19200 Bd
Figure 45: Selecting the baud rate for the RS232 interface
Setting of the baud rate can also be carried out using the DataVIEW or
DataEXPERT communication module via PC.
In order to avoid disturbances with the AlphaGUARD monitor, if possible please
use the delivered RS
-
232/USB cable. Longer cables can cause disturbances
due to EMI interferences and thus malfunctions of the instrument. For on-
operation over longer distances with RS-232
cable the AlphaGUARD interface
must be galvanically separated from the cable connection.
Only use interface modules with an own power supply.